Customization: | Available |
---|---|
Type: | Anaesthesia Machine |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
HA-3800 Trolley Anesthesia Ventilator Machine
This HA-3800 Anesthesia Machine is equipped with an accurate dedicated anaesthetic vaporizer and a safety device for preventing cyanosis and necessary alarm system. During the anesthesia, the patient's respiratory functions can be managed by using a microcomputer controlled pneumatic electrically controlled synchronizing anesthesia respirator. Each connection part of the whole machine is a standard interface. A highly efficient and large volume soda lime absorber can minimize the re-inhalation of carbon dioxide by the patient.
Power On Self Test
The automatic conversion for standby power while AC power outage, keep working not less than 1 hour.
Promoting Excellence Through Design
Built from nonhazardous environmentally friendly material, the precisely constructed and elegantly designed Jinling 850 Adv. provides an ergonomic and easily movable anesthesia machine that requires only the bare minimum of operating space.
Physical specifications
|
|
Screen
|
8.4 / 10.4 inch LCD Display Screen
|
Suitable
|
Adult & Child
|
Mode
|
pneumatically powered and electrically controlled system
|
Working Mode
|
Closed; Semi-Closed; Semi-Open
|
Circuit
|
Breathing circuit integrated standards
|
Flowmeters
|
5 Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min; Air: 0.1~10L / Min
|
Trolley
|
Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
|
Drawer unit
|
TWO drawers totally extractable
|
Technical Specifications
|
|
Gas requirement
|
Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa; NO2: 0.32 MPa to 0.6 MPa.
|
safety valve
|
<12.5 kPa
|
Respiratory Rate
|
1~99bpm
|
oxygen concentration in the mixed gas N2O/O2
|
> 21%
|
Oxygen Flush
|
25~75 L/min
|
Modes of ventilation
|
A/C, IPPV, SIPPV, IMV, SIMV, PCV, VCV, PEEP, SIGH, PEEP, MANUAL
|
PEEP
|
0 ~ 20 hPa
|
Ptr
|
-10 ~ 20 hPa
|
IMV frequency
|
1 ~ 20 bpm
|
I/E Ratio
|
8:1 ~ 1:10, Has the inverse ratio ventilation
|
Tidal Volume
|
0 ~ 1500 ML
|
Inspiratory Plateau
|
0~1s
|
O2 Concentration
|
21%~100%
|
Sigh
|
one deep breath per 70 ~ 120 controlled respirations, the inspiration time is 1.5 times of the setting point
|
Maximum safety pressure
|
≤ 12.5 kPa
|
Pressure limit range
|
0 ~ 6.0 kPa
|
Airway pressure alarm: Audible and visual and with yellow and red color indicating
|
Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
|
±0.2 kPa
|
|
Tidal volume alarm: Audible and visual and with yellow and red color indicating
|
upper alarm: 50 to 2000ml, lower alarm: 0~1800ml
|
Oxygen concentration alarm: Audible and visual and with yellow and red color indicating
|
upper alarm: 21% ~100%; lower alarm:10%~80%
|
Power Supply Alarm
|
Ac/dc power supply are after failing to send out alarm immediately Alarm Time: keep >120s |
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
|
|
Safty & Alarm
|
|
Oxygen Alarm
|
It alarms when the oxygen supply from pipe or cylinders lower than 0.2MPa
|
Ventilation Volume Alarm
|
Lower: 0~12L/Min; Up: 18L/ Min
|
Power Alarm
|
It Alrmas while AC and DC supply failure; Keep alraming time: >120s
|
Air Tract Pressure Alarm
|
Lower: 0.2kPa ~5.0 kPa; Up: 0.3kPa ~ 6.0kPa
|
1. What about the price level?