Technical Specifications | |
ECG | |
Lead mode | 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V |
Protection | Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof |
Gain | 2.5mm/mV(×0.25), 5.0mm/mV(×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto |
Sweep speed | 6.25mm/s, 12.5mm/s, 25mm/s, 50 mm/s |
ECG signal range | ±12 mV (Gain × 0.25) |
Accuracy | ±1bpm/ ±1%, whichever is greater |
Resolution | 1 bpm |
Leakage Current | < 10 μA |
Baseline Recovery | ≤ 3s after defibrillation (Monitor mode) ≤ 1s after defibrillation (Surgery mode) |
Bandwidth | Surgery 1 ~ 25 Hz Monitor 0.5 ~ 40 Hz Diagnostic 0.05 ~ 120 Hz |
Indication of Electrode Separation | Every electrode (exclusive of RL) |
Heart Rate | |
Measure range | Adult: 0bpm; 10 ~ 300 bpm Neo/Ped: 0bpm; 10 ~ 350 bpm |
Resolution | 1 bpm |
Accuracy | ± 1% or ± 1bpm, whichever is greater |
ST Measurement | |
Range | -2.0 ~ +2.0 mV |
Accuracy | -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is greater |
Other range | unspecified |
Resolution | 0.01mV |
Respiration | |
Method | Impedance between RA-LL, RA-LA |
Gain | ×0.25, ×0.50, ×1, ×2, ×4 |
Respiration Rate | 0bpm, 6 ~ 150 BrPM |
Sweep speed | 6.25 mm/s, 12.5 mm/s, 25mm/s |
Resolution | 1 BrPM |
Accuracy | ±2 bpm or ±2%, whichever is greater |
Apnea Alarm | 10 ~ 60 s |
NIBP | |
Method | Oscillometric |
Measure mode | Manual, Auto, STAT |
Measure Interval in AUTO Mode | 1~480 min |
STAT mode cycle time | Keep 5 minutes, at 5 seconds interval |
Measure and Alarm Range | Adult: SYS: 30 ~ 270 mmHg DIA: 10 ~ 220 mmHg MEAN: 20 ~ 235 mmHg Pediatric: SYS: 40 ~ 235 mmHg DIA: 10 ~ 220 mmHg MEAN: 20 ~ 225 mmHg Neonate: SYS: 30 ~ 135 mmHg DIA: 10 ~ 110 mmHg MEAN: 20 ~ 125 mmHg |
Static pressure accuracy | ±3mmHg |
Resolution | 1mmHg |
Accuracy | Maximum Mean error ±5mmHg Maximum Standard deviation ≤8mmHg |
Over pressure Protection | Dual protection via software & hardware |
Temperature | |
Technique | Thermistor probe (2.25K) |
Channel | Dual-channel, provide T1; T2; ΔT |
Measuring and Alarm Range | 0.0 ºC~ 50 ºC (32ºF~ 122ºF) |
Unit | Celsius (ºC), Fahrenheit ( ºF) |
Resolution | 0.1ºCor 1ºF |
Accuracy | ±0.1ºC(exclusive probe) |
Pulse Rate | |
Measurement range | 20 ~ 300 bpm |
Resolution | 1bpm |
Accuracy | ±3 bpm (20 ~ 250 bpm) unspecified (251~300 bpm) |
1. What about the price level?
We are confident that our prices are competitive compared with those at the same quality level, since we have stayed in this filed more than 20 years. And we always adhere to the principle of provide top quality products with competitive prices. We have already established our good fame in many countries, especially in Africa, Middle East, European, Asia and South American through year after year efforts.
2. Why we find some prices are very high compared with some other supplier, and from the picture or it's model number, all looks the same?
As China is still a developing country, there are only a few products with patent rights, and copy products I spread very fast in the market. Of course, the prices of those copy products with cheap material and less quality control are relatively cheaper.
3. What is time of delivery?
Since our products are produced according to each specific order, the delivery period will be 15 days after confirmation of order. If it is small or sample order, please do check with us about whether we have ready goods in stock to make delivery as soon as possible.
4. What about quality assurance?
We offer one year warranty period beginning from the time of shipment for medical equipment, Per minor problems, we can provide free spare parts for replacements, Per serious problems, we can make replacement of free charge.
5. What is payment terms?
T/T, Credit Card, L/C, Western Union, PayPal, MoneyGram.