Service Features
1. The Support Tool helps technical personnel to
- carry out configuration, upgrades and troubleshooting on an individual monitor.
- back up the monitor settings.
2. The Service Mode is password-protected and ensures that only trained staff can access service tests and tasks. It includes Printer Setting, Skin Type, AFM Setting, Brightness Setting, Clear Data, NIBP Calibration, and Touch Screen Calibration.
3. The Support Tool uses the USB interface of the fetal monitor for software upgrading
Performance Specifications
Dimension and Weight
- Dimension: 360mm*310mm*120mm
- Weight: 5.3kg(excluding accessories)
Power Supply
- Voltage: AC100~240V, 50/60HZ
Display
- 12.1" color TFT LED (0-90° rotation)
- Resolution: 1024*768 pixels
Battery (Built-in)
- Type: Rechargeable lithium battery 14.8V/2200mAh
- Charge Cycle: ≥500 times
- Working time: 2 hours
Recorder (Built-in)
- Method: Thermal printer
- Paper: Z-fold, thermosensitive
- Paper width: 112 mm (4.41 in)/150 mm (5.91 in)
- Printing speed: 1/2/3cm/min
- Trace: Max. 4 tracks
- Recording way: Real-time Recording, Review Printing
Alarm
- Level: Low, medium and high
- Indication: Auditory, Visual, Delay of FHR Alarm
Input Device
- Touch screen (option)
- Knob (Standard)
- Keypad input (Standard)
- Mouse/ Keyboard input (option)
- USB port for patient data/NST report backup
(option)
System Output & Extensible
Interface
- Ethernet Network: 1 standard RJ45 socket
- USB Port: 2
- Video Output: 1 HDMI port (option)
Trend & Reviewing
- Cardiotocography (CTG):
Waveform Review: 8640 hours
- Vital Signs:
Trend: 2160 hours
Waveform Review: 2 hours
NIBP Review: 2000 groups
- Alarm Review: 2000 groups
Safety
- IEC60601-1 Approved, CE marking according to MDD93/42/EEC
- With reference to RoHS Directive 2011/65/EU recasting
Ultrasound (FHR)
- Transducer type: 12-Crystal ultrasound transducer
- Measure method: Wide beam pulse doppler
- FHR Measurement range:
30 to 240 bpm (US Standard)
50 to 210 bpm (Europe Standard)
- Accuracy: ±1 bpm
- Intensity: <3 mW/cm2
- Ultrasound frequency: 1MHz±3%
- Resolution: 1 bpm
TOCO
- Measurement method: Strain gauge sensor element
- Measure range: 0 ~ 100 units
- Resolution: 1 unit
- Nonlinear deviation : ≤±8%
- Baseline setting: 0,5,10,15, 20 can be selected
- Auto offset correction: 3 second after connecting the transducer
- Auto zero adjust: TOCO value is set to zero negative following a negative Measure value for 5 seconds
Fetal Movement Recording
- Automatic fetal movement (AFM): pulsed doppler ultrasound
- Automatic fetal movement autograph
- Manual fetal movement marker (MFM)
Maternal ECG
- Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
- Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, Auto
- Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50 mm/s
- Accuracy: ±1bpm/ ±1%, whichever is greater
- Resolution: 1 bpm
- Leakage Current < 10 μA
- CMRR: Diagnostic Mode >89 dB
Monitor Mode >105 dB
Surgery Mode >105 dB
- Bandwidth: Surgery 1 ~ 25 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 120 Hz
Maternal Heart Rate
- Measure range: Adult: 0bpm; 10 ~ 300 bpm
Neonate: 0bpm; 10 ~ 350 bpm
- Resolution: 1 bpm
- Accuracy: ± 1% or ± 1bpm, whichever is greater
ST Measurement
- Range: -2.0 ~ +2.0 mV
- Accuracy: -0.8mV ~ +0.8mV: ±0.02mV or ±10%, whichever is greater
- Other range: unspecified
- Resolution: 0.01mV
Maternal Respiration
- Method: Impedance between RA-LL, RA-LA
- Gain: ×0.25, ×0.50, ×1, ×2, ×4
- Respiration Rate: 0bpm, 6 ~ 150 bpm
- Resolution: 1 bpm
- Accuracy: ±2 bpm or ±2%, whichever is greater
- Apnea Alarm: 10 ~ 60 s
Maternal SpO2 (Digital Technic)
- Measurement Range: 0 ~ 100 %
- Resolution: 1 %
- Accuracy: ±2% (70% ~ 100%)
±3 % (35% ~ 69%)
Unspecified (0 ~ 34%)
Pulse Rate
- Range: 25~250bpm
- Accuracy: ±1% or ±1 bpm, whichever is greater
- Resolution: 1bpm
Maternal NIBP (MNIBP)
- Method: Oscillometric
- Measure mode Manual, Auto, STAT
- Measure Interval in AUTO Mode 1~480 min
- STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
- Measure and Alarm Range:
Systolic: 30~270 mmHg
Diastolic: 10~220 mmHg
Mean: 20~235 mmHg
- Static pressure accuracy ±3mmHg
- Resolution: 1mmHg
- Accuracy: Maximum Mean error ±5mmHg
Maximum Standard deviation 8mmHg
- Over pressure Protection: Dual protection via software & hardware
Maternal Temperature
- Technique: Thermistor probe (2.25K)
- Measuring and Alarm Range: 0.0 ºC~ 50 ºC (32ºF~ 122ºF)
- Unit: Celsius (ºC), Fahrenheit ( ºF)
- Resolution: 0.1ºC
- Accuracy: ±0.1ºC
NST Report/FHR Interpretation
(Option)
- Based on Fischer Evaluation Method (monitoring time ≥20 Min)
- Average Baseline FHR / Number of Acceleration / Number of Deceleration / Number of Small Acceleration / Amplitude Variation / Number of PV
- Automatic Score Result: Sore: 8-10 (Excellent)
Sore: 5-7 (Suspicious)
Score:≤4 ( Fetus Hypoxia)