Gold Member Since 2011
Audited Supplier
Healicom Medical Equipment Co., Limited

Medical Equipment, Incubator, baby incubator manufacturer / supplier in China, offering Medical Equipment Ce Approved Infant Incubator H-2000, HL-05 Vertical Luminescence Shadowless Lamp, Hdc-99b-II Medical Gynecological Operation Table and so on.

(/ )

Supplier Homepage Product Infant Equipments Infant Incubator Medical Equipment Ce Approved Infant Incubator H-2000

Medical Equipment Ce Approved Infant Incubator H-2000

FOB Price: US $1,500 / Piece
Min. Order: 1 Piece
Min. Order FOB Price
1 Piece US $1,500/ Piece
Get Latest Price
Port: Shanghai, China
Production Capacity: 1000units Per Month
Payment Terms: T/T, Western Union

Request a custom order and have something just for you!

Send Customized Request
Basic Info
  • Model NO.: H-2000
  • Medical Device Regulatory Type: Type 3
  • Temperature Rise Time: ≤30min
  • Infant Bed Max Load: 10kg
  • Warranty: 12 Months
  • Transport Package: Carton
  • Origin: Jiangsu, China
  • Medical Devices Reg./Record No.: Gsyjxzz20123540587
  • Power Input: ≤400va
  • Cabinet Max Load: 6kg
  • Certificate: CE, ISO
  • Trademark: Healicom
  • Specification: CE, ISO
  • HS Code: 9018909090
Product Description
Standard Composition
  1 Servo Control Mode: air temp, skin temp;
  2 Instant Parameter Displaying: set temp, air temp ,skin temp, heater power rate;
  3 Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin temp override, fan failure, system failure;
  4 >37°C Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
  5 Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;
  6 RS-232 Output Connector (Use for network system)
 
Specification
PowerModel H-2000-1E110 / 120 V ac, 50Hz / 60Hz
Model H-2000-2E220 / 240 V ac, 50Hz / 60Hz
Power Input≤400VA
Refer to IEC60601-1, this equipment belongs to ClassI, Type BF 
Its operation is continuous
AlarmPower AlarmPower Failure: Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle
Fan Alarm Fan Failure: Activates if Fan motor malfunctions
Sensor AlarmAir Sensor Failure: Air Sensor malfunction, refer unit to service
Sensor Out of Position: Activated when Sensor is not in correct position to perform calibration or continue operation
Skin Sensor Disconnected: Activated during Skin Control Mode when Skin Sensor is disconnected from Temperature Controller
Skin Sensor Failure: Activated when Skin Sensor malfunctions
Deviation AlarmHigh Air Temperature: Activated if displayed Air Temperature is 15°C above set point
Low Air Temperature: Activated if displayed Air Temperature is 25°C below set point
High Skin Temperature: Activated if displayed Skin Temperature is 10°C above set point
Low Skin Temperature: Activated if displayed Skin Temperature is 10°C below set point
Over Temperature AlarmOver Air Temperature: Activated when displayed Air Temperature was approach 38°C for set temperature ≤37°C, or approach 40°C for set temperature >37°C in Air Mode
Over Skin Temperature: Activated when displayed Skin Temperature was approach 38°C for set temperature ≤37°C, or approach 39°C for set temperature >37°C in Skin Mode
System AlarmSystem Failure: Activated if Temperature Control System malfunctions
Temperature Control Range and Correlative SpecificationsAir Temperature20°C~37°C,for set temperature ≤37°C, 37°C~39°C, for set temperature >37°C
Skin Temperature20°C~37°C,for set temperature ≤37°C, 37°C~38°C, for set temperature >37°C
Temperature Rise Time≤30min
Temperature Variability±0.5°C
Temperature Uniformity≤1°C
Skin Sensor Precision≤0.3°C
Environment RequirementEnvironment TemperatureOperation Temperature Range+20°C~+30°C
Store Temperature Range-40°C~+55°C
Environment HumidityOperation Humidity Range30%~75%
Store Humidity Range10%~100%
Atmospheric Pressure Range700hPa~1060hPa
Other
Specifications
Noise Level Within Hood Environment≤55dB(Environment noise below 45dB)
Carbon Dioxide (CO2) Level within the HoodLess than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a point 10cm above the center of the mattress
Air Velocity Over Mattressless than 0.3m/s
Infant Bed Max Load10kg
Transfusion Shelf Max Load2kg
Cabinet Max Load6kg
Function of "SILENCE" KeyMake the alarm to be silence for 10 minutes by press this keyDeviation Alarm
Sensor Alarm
Fan Alarm
Factory Default SettingsTemperature Control Mode Air Temperature Mode
Air Set Temperature: 32°C
Skin Set Temperature: 34°C
Temperature Units:°C
      
 
NOTE: Open Access Panel or Access Door or the use of supplies or other equipment within the Incubator, which can alter the air flow pattern, may effect temperature uniformity, temperature variability, the correlation of the Incubator temperature reading to center mattress temperature and skin temperature

Medical Equipment Ce Approved Infant Incubator H-2000
Medical Equipment Ce Approved Infant Incubator H-2000Medical Equipment Ce Approved Infant Incubator H-2000Medical Equipment Ce Approved Infant Incubator H-2000
 
Send your message to this supplier
Avatar
Miss Mary
*From:
To: Miss Mary
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now